- Institutional procedure manuals should establish working conditions according to the load, volume, weight and thermal resistance of the material.
- It is indispensable to respect the parameters obtained during the validation of the procedure.
- Temperature: the temperature of sterilization by dry heat should stay between 160 ºC – 170 ºC.
- Times: the total exposure time of the material is determined through the corresponding validation of the cycle.
- It is important to point out that the exposure time should be recorded after the required temperature is reached and not from the time that the sterilizer is charged since a prolonged time could be required to reach the sterilization temperature.
Relationship between time – temperature for sterilization by dry heat